FD-92
Calibration of FD-92
Summary:
A step-by-step tutorial on calibrating your FD-92 gas detector.
Model:
FD-92 all versions including FD-92-NH3, FD-92-LEAK, FD-92-FREON
Calibration Frequency:
At least every 12 months.
Calibration Kit Items Required:
-
Calibration Gas Bottle
- Select the corresponding calibration gas for your unique gas detector. To do so, go to the product listing of your gas detector for the recommended calibration gas bottle.
- Calibration Gas Regulator (C10)
- Tubing
- FD-92 Calibration Cap
![]() |
▶️ Tutorial Calibration Video Example: FD-92-NH3
|
Note: Example of Ammonia (FD-92-NH3). Each FD-92 is unique and made to a specific single target gas. Follow this procedure using the recommended calibration gas bottle for your particular target gas.
✅ STEP 1: ZERO Calibration to 0 ppm using Fresh Air.
- Power on your gas detector and allow 2 minutes for stabilization in fresh air.
- Expose the gas analyzer to fresh air for at least 2 minutes.
- The gas detector may or may not read precisely 0 ppm. Depending on the model, it may or may not read 0.0 or 0.00 ppm. To zero calibrate and make it read exactly zero, undertake the following steps.
- Press the middle OK button to go to the Menu Screen.
- Select the Calibration option.
- Enter password 11 and the logn press the middle OK button.
- Select "Cal. Zero" by pressing the middle OK button.
- Once done you will see "Success"
- Done!
- In the main display screen, the unit will now read 0 ppm in fresh air.
✅ STEP 2: SPAN Calibration.
- Ensure you first undertake a Zero Calibration.
- Then, expose the gas analyzer to your calibration reference gas for 2 minutes or until the reading has stabilized. Use a flow rate of 0.5 LPM and connect the gas to the gas detector using the FD-92 calibration cap.
- Press the middle OK button to go to the Menu Screen.
- Select the Calibration option.
- Enter password 11 and press the middle OK button.
- Go down to highlight the Calibration Value and change the value to be the same as your reference gas.
- Enter the numeric value via the key pad then press green tick.
- Move the highlighted orange box to the Calibration box and press OK.
- Once done, you will see "Success"
- Go back to the main display and confirm the digit value is the same as your reference gas. Done!
✅ STEP 3: Verification
- Allow the gas detector to stabilize in fresh air for 5 minutes, confirming it reads 0 ppm in fresh air.
- Expose the analyzer to the calibration reference gas to verify the calibration and detection repeatability.
- If the gas detector reading is within +/- 2% F.S. of the calibration gas value, this is considered a PASS. For example, FD-90A-CO has a full scale of 1000 ppm. The unit was calibrated to a reference gas of 200 ppm. Then, a PASS range is +/-2% of 1000 ppm = +/- 20 ppm. Therefore, a PASS is between 180 ppm and 220 ppm.
- Once verified, your unit is properly calibrated and ready for operation.
- Don't forget to document the calibration event in your gas analyzer log book. Place a reminder calibration date sticker on the unit for the next calibration event.

